西门子SPECT(E.CAM)仪“U—FLY”均匀性校正应用 预览 被引量:3

Application of uniformity correction technology of Siemens E.CAM U-fly
在线阅读 下载PDF
收藏 分享 导出
摘要 通过对siemens“U-fkt”技术的分析与实践,评价其对仪器系统的作用。运行系统配置的基本测试程序,分别采用^99mTc和^57Co标准源.依指令顺序操作,对探头的非均匀性可进行有效校正,取得了满意结果。由于均匀性指标是影响临床图像质量最重要的一个因素,建议每日采集图像前进行该项测试。 The efficiency of Siemens E.CAM is evaluated by analyzing and practicing the technology of Siemens U-fly. Basic testing procedure is run. Standard source ^99mTc and ^57Co is used respectively and the operation is performed according to the instruction. The detecting bead's uniformity can be corrected in this way with satisfactory results. It is suggested uni- formity should be tested every day before study because it is the most important factor for the quality of clinical images.
作者 张书文 何义杰 邵明哲 ZHANG Shu-wen, HE Yi-jie, SHAO Ming-zhe (General Hospital of PLA, Beijing 100853, China)
机构地区 解放军总医院
出处 《医疗卫生装备》 2006年第9期 70,73,共2页 Chinese Medical Equipment Journal
关键词 仪器 图像质量 均匀性 校正技术 instrument image quality uniformity correction technology
  • 相关文献

参考文献3

  • 1NEMA standards publication NU 2-1994.National Electrical Manufactures Association,1994 被引量:1
  • 2Hines H,Kayayan R,Colsher J,et al.National Electrical Manufactures Association Recommendations for Implementing SPECT Instrumentation Quality Control.J Nucl Med,2000,41 (2):383~389 被引量:1
  • 3Kuikka JT,Yang J,Kiiliainen H.Physical performance of the Siemens E.CAM gamma camera.Nucl Med Commun,1998,19(5):457~462 被引量:1