比较两种低剂量CT迭代重建法与常规剂量滤波反投影重建法的图像质量 预览 被引量:18

Comparison on image quality of two kinds of low-dose CT reconstruction and routine-dose CT filtered back projection reconstruction
在线阅读 下载PDF
收藏 分享 导出
摘要 目的对比低剂量腹部和盆腔增强扫描自适应统计迭代重建(ASIR)和基于模型的迭代重建(MBIR)与常规剂量下传统的滤波反投影(FBP)重建的图像质量和剂量减低率。方法31例患者接受腹部和盆腔增强常规剂量CT扫描、FBP重建;复查时,接受增强低剂量扫描、40%ASIR和MBIR。由2名医师通过锐利度、噪声、伪影和诊断接受度对图像进行评分,并测量噪声值和CT值,计算SNR。记录每例患者每次检查的剂量长度乘积(DI。P)和CT剂量指数(CTDIv—01),计算剂量减低率。结果低剂量扫描的DI.P值和CTDI值分别为(328.95±206.35)mGy/cm和(7.96±4.30)mGv,而常规剂量FBP重建的DLP值和CTDl分别为(689.27±339.63)mGy/cm和(16.81±7.19)mGy。对于腹部和盆腔脏器,低剂量MBIR图像比低剂量40%ASIR图像和常规剂量FBP图像有更低噪声值和更高SNR(P均〈0.0167)。低剂量40%ASIR图像和常规剂量FBP图像的客观评价结果相似(P〉0.0167)。MBIR和40%ASIR可以提高图像的密度分辨力,降低硬化伪影,且MBIR比40%ASIR图像有更低噪声和伪影,主观评价结果更佳。结论在保证图像质量的前提下,相比常规剂量扫描、FBP重建,低剂量扫描MBIR和ASIR可以明显降低扫描剂量;相比ASIR法,MBIR法能提供更佳图像质量,具有进一步降低扫描剂量的潜力。 Objective To evaluate dose reduction and image quality characteristics of abdominal low-dose CT reconstruc- ted with model-based iterative reconstruction (MBIR) and adaptive statistical interactive reconstruction (ASIR) compared with routine-dose CT reconstructed and filtered back projection (FBP). Methods Thirty-one patients underwent contrast- enhanced abdominal-pelvis low-dose CT reconstructed with 40~ASIR and MBIR after contrast enhanced abdominal-pelvis routine-dose CT reconstructed with FBP. Two radiologists assessed the images blindly according to the sharpness, image noise, diagnostic acceptability and artifacts. Image noise and SNR of abdominal organs relative to abdominal fat were as- sessed. The volume CT dose index (CTDIvol), dose-length product (DLP) and dose reduction rate were also obtained. Re- sults CTDIvol and DLP for low-dose CT were (7.96- 4.30)mGy and (328.95 ~ 206.35)mGy/cm, of routine-dose CT with FBP were (16.81 ~ 7.19)mGy and (689.27 i-339.63)mGy/em, respectively. Low-dose MBIR images had significant- ly lower objective image noise, higher SNR than low-dose 40~ASIR images and routine-dose images (all P%0. 0167). The low-dose 40~ASIR and routine-dose images had similar objective and subjective results (all P〉0. 0167). MBIR had better subjective scores than 40~/00ASIR and FBP in noise, sharpness and artifacts. Meanwhile, MBIR and ASIR could not only improve the density resolution, but also c decrease the hardening artifacts of images. Conclusion Under the premise of same image quality, low-dose CT reconstruction can decrease radiation dose significantly compared to routine-dose CT re- construction. Image quality of low-dose CT MBIR is better than that of 40~ASIR, which has potential for further reducing radiation dose.
作者 杨晶 高艳 李坤成 杜祥颖 李岩 YANG J ing , GAO Yan , LI Kun-cheng , DU Xiang ying , LI Yan (Department of Radiology, Xuanwu Hospital, Capital Medical University, Beijing 100053, China)
出处 《中国医学影像技术》 CSCD 北大核心 2014年第1期117-121,共5页 Chinese Journal of Medical Imaging Technology
关键词 体层摄影术 X线计算机 低剂量 重建 Tomography X-ray computed Low dose Reconstruction
  • 相关文献

参考文献13

  • 1Schauer I)A, l.inton ()W. N('RP report no. 160, ionizing radia lion exposure of the population of the United States, medical ex- posure-are we doing less with more. amt is there a role for health physicists? Health Phys, 2009.97(1 ) : 1-5. 被引量:1
  • 2Brenner DJ, Hall EJ. Computed tomography- an increasing source of radiation exposure, N Engl J Med, 2007, 357 (22): 2277-2284. 被引量:1
  • 3宁培钢,朱绍成,史大鹏,郭濴,孙明华.高级迭代重建算法降低腹部CT剂量的潜能:体模研究[J].中国医学影像技术,2012,28(12):2243-2247. 被引量:18
  • 4王艳,史大鹏,朱绍成,吴青霞,孙明华.比较以自适应性统计迭代重建技术和滤过反投影重建的低剂量腹部CT的图像质量[J].中国医学影像技术,2012,28(10):1902-1905. 被引量:35
  • 5Sagara Y, Hara AK, Pavlicek W, et al. Abdominal CT: Com- parison ot" low dose CT with adaptive statistical iterative Recon- struction and routine dose CT with filtered back projection in 53 patients. AJR AmJ Roentgenol, 2010, 195(3):713-719. 被引量:1
  • 6Akash P, Kalra MK, Digum*trthy SR, et al. Radiation dose re- duction with chest computed tomography using adaptive starts ti- cal iterltive Reconstruction technique: Initial experience. J Com- put Assist Tomogr, 2010,34(1):,10 45. 被引量:1
  • 7l'rakash I, Kalra MK. Ackman Jl'l, et al. Diffuse lung disease: C'F of the chest with adaptive statistical iterative recon-struction lecbnique. Radiology, 2010,256( 1 ) :9'61-269. 被引量:1
  • 8Boedeker KI., McNiu Gray MF. Application of the noise power speelrum in modern diagnoslic MDCT: Part H- Noise power spectra and signal to noise. Phys Med Biol, 2007,52(1,1):40,17 ,:1061. 被引量:1
  • 9Hara AK, Paden R(;, Silva AC, et al. herative reconstruction technique for reducing body radiation close at CT: Feasibilitystudy. AJR Am J Roentgenol, 2009, 193(3):764-771. 被引量:1
  • 10Katsura M, Matsuda I, Akahane M, et al. Model-based itera- rive Reconstruction technique for radiation dose reduction in chest (7"I": Comparison with the adaptive statistical iterative Re- construclion technique. Eur Radiol, 2012.22 (8) : 1613 1623. 被引量:1

二级参考文献24

  • 1EUR 16262. European guidelines on quality criteria for computed tomography, www. drs. dk/guidelines/et/quality/download/ cur16262.wS]. Accessed September 10,2009. 被引量:1
  • 2Singh S. Kalra MK. Moore MA, et al. Dose reduction and com- plication with pediatric CT protocols adapted to patient size, clini- cal indication and number of prior studies. Radiology, 2009,252 (l) :200-208. 被引量:1
  • 3Lee CI, Haims AH, Monico EP, et al. Diagnostic CT scans: As- sessment of patient, physician, and radiologist awareness of radi- ation dose and possible risks. Radiology, 2004,231(2):393-398. 被引量:1
  • 4Thibault JB, Sauer KD, Bouman CA, et al. A three dimensional statistical approach to improved image quality for multislice helical CT. Med Phys, 2007,34(11) :4526-454,4. 被引量:1
  • 5Hara AK, Paden RG, Silva AC, et al. Iterative Reconstruction technique for reducing body radiation dose at CT: Feasibility study. AJR Am J Roentgenol, 2009,193(3):764-771. 被引量:1
  • 6Prakash P, Kalra MK, Kambadakone AK, et al. Reducing ab- dominal CT radiation dose with adaptive statistical iterative recon- struction technique. Invest RadioI, 2010,45 (4) : 202-210. 被引量:1
  • 7Sagara Y, Hara AK, Pavlicek W, et al. Abdominal CT: Com- parison of low-dose CT with adaptive statistical iterative recon- struction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol, 2010, 195(3) :713-719. 被引量:1
  • 8Ren Q, Dcwan SK, Li M, et al. Comparison of adaptive statisti- cal iterative and filtered back projection reconstruction techniques in brainCT. Eur J Radiol, 2012,11(9) :5842-5847. 被引量:1
  • 9Yanagawa M, Honda O, Kikuyama A, et al. Pulmonary nod- ules: Effect of adaptive statistical iterative reconstruction (ASIR) technique on performance of a computer-aided detection (CAD) system--Comparison of performance between different-dose CT scans. EurJ Radiol, 2011,41(12):5693-5703. 被引量:1
  • 10Scheffel H, Stolzmann P, Schlett CL, et al. Coronary artery plaques : Cardiac CT with model-based and adaptive-statistical it- erative reconstruction technique. Eur J Radiol, 2012, 81 (3): e363-e369. 被引量:1

共引文献48

同被引文献134

引证文献18

二级引证文献23

投稿分析
职称考试

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部 意见反馈