不同采集矩阵与计数对SPECT静态显像的影响 预览 被引量:1

Influence of matrix and counts on gamma camera static imaging
在线阅读 下载PDF
收藏 分享 导出
摘要 目的:研究不同采集矩阵与计数对单光子计算机断层(SPECT)静态显像的影响。方法:通过SPECT/PET NEMA 2001模型,研究不同采集矩阵与不同计数状态下的图像质量,对模型图像的噪声和对比度进行分析。使用线源测定不同采集矩阵的系统分辨率。结果:随着计数的增加,图像的噪声减少;低计数状态下小矩阵图像的对比度优于大矩阵图像;高计数状态下大矩阵图像的对比度优于小矩阵图像。128×128矩阵的半高宽(FWHM)为(10.52±0.86)mm,256×256矩阵的FWHM为(9.47±0.61)mm,512×512矩阵的FWHM为(8.79±0.97)mm。结论:SPECT静态显像时低计数状态下采用小矩阵可以获得相对较高质量的图像,高计数状态下采用大矩阵可以获得相对更高质量的图像;随着采集矩阵的增大,系统分辨率略有提高。 Objective:To investigate the influence of different acquisition matrix and counts on Gamma camera static imaging.Methods: 1)The image quality under different acquisition matrix and counts was studied on the SPECT/PET NEMA 2001 model, we analyzed the influence of different acquisition matrix and counts on the noise and contrast. 2)We use the linear source to determine the system resolution in different acquisition matrix.Results: 1)With each increase in acquisition counts, the noise will decrease; The use of small matrix and low counts can get low noise images; and the use of large matrix and high counts can get high spatial resolution images. 2)The FWHM(mm) with 128×128, 256×256, 512×512 matrix were respectively 10.52±0.86, 9.47±0.61, 8.79±0.97.Conclusion: In a static collection process, the use of small matrix and low counts can get high quality images and the use of large matrix and high counts can get better effectiveness of images; with each increase in the matrix, the system resolution will increase slightly.
作者 王猛 金超岭 郑玉民 颜珏 WANG Meng, JIN Chao-ling, ZHENG Yu-min, et al (Department of Nuclear Medicine, (Jhina-Japan Friendship Hospital, Beijing 100029, China.)
出处 《中国医学装备》 2015年第9期50-53,共4页 China Medical Equipment
关键词 矩阵 计数 伽马照相机 静态显像 Matrix Counts Gammacamera Static imaging
  • 相关文献

参考文献16

二级参考文献78

共引文献54

同被引文献13

  • 1Caobelli F,Pizzocaro C,Paghera B,et al.Evaluation of patients with coronary arterydisease.IQ-SPECT protocol in myocardialperfusion imaging:Preliminary results[J].Nuklearmedizin,2013,52(5):178-185. 被引量:1
  • 2Garcia EV,Faber TL,Esteves FP.Cardiac dedicatedultrafast SPECT cameras:new designs and clinicalimplications[J].J Nucl Med,2011,52(2):210-217. 被引量:1
  • 3Imbert L,Poussier S,Franken PR,et al.ComparedPerformance of High-Sensitivity CamerasDedicated to Myocardial Perfusion SPECT:AComprehensive Analysis of Phantom and HumanImages[J].J Nucl Med,2012,53(12):1897-1903. 被引量:1
  • 4Gharib H,Papini E,Paschke R,et al.AmericanAssociation of Clinical Endocrinologists,AssociazioneMedici Endocrinologici,and European ThyroidAssociation;Medical guidelines for clinical practicefor the diagnosis and management of thyroidnodules[J].Endocr Pract,2010,16(Suppl 1):1-40. 被引量:1
  • 5Murazaki H,Funama Y,Sugaya Y,et al.OptimalSetting of Automatic Exposure Control Based onImage Noise and Contrast on Iodine-enhancedCT[J].Acad Radiol,2012,19(4):478-484. 被引量:1
  • 6Koulouri O,Moran C,Halsall D,et al.Pitfalls inthe measurement and interpretation of thyroidfunction tests[J].Best Pract Res Clin EndocrinolMetab,2013,27(6):745-62. 被引量:1
  • 7Jackson J,Pan T,Tonkopi E,et al.Implementationof automated tube current modulation in PET/CT:prospective selection of a noise index andretrospective patient analysis to ensure imagequality[J].J Nucl Med Technol,2011,39(2):83-90. 被引量:1
  • 8张秀梅,方凤宁,王亚平,冯珏.SPECT/PET系统模型试验技术探讨[J].中国医疗设备,2010(5):24-26. 被引量:2
  • 9王渊恺,张光明,韩芳,刘从进,刘兴党.不同采集矩阵与函数重建对SPECT影像质量影响[J].中国医学计算机成像杂志,2011,17(2):164-167. 被引量:7
  • 10任志刚,李小华.SPECT的数据采集条件对其断层分辨率的影响实验探讨[J].中国医学物理学杂志,2000,17(1):4-5. 被引量:5

引证文献1

二级引证文献3

投稿分析

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部 意见反馈